CONCLUSION With the current technical status, speech recognition using cochlear implants is good in quiet surroundings and there is even room left for speech recognition in background noise. With the use of directional microphone systems, this speech recognition in background noise can be substantially improved. Compared to an average speech reception threshold of +2.5 dB found with the standard headpiece microphone of the CII cochlear implant, the Handymic resulted in a benefit of 8.2 dB and the Linkit array microphone system in a benefit of 5.9 dB. When both directional microphone systems were tested, the CI subjects were able to recognize more than 62% of the phonemes presented at 0 dB SNR. This might be of great importance in situations such as restaurant or cocktail party settings. The improvement could make a difference in the way communication can be carried out. Instead of guessing the line of conversation by listening and lip-reading, this benefit could result in a fair intelligibility and so be of significant help in understanding what is said and in easing conversation.