Proefschrift Kerklaan
Chapter 3
Table 1b. Ventilator settings and metabolic measurement data for all children and for subgroups based on compatibility of the indirect calorimetry (IC) and ventilator method (defined as a difference ≤10% or >10% between IC and ventilator-derived VCO 2 values) All children P-value
Difference ≤ 10% N=18
Difference >10% N=23
N=41
Ventilation mode Support Control
%
0.94 c
34.1 65.9
33.3 66.7
34.8 65.2
FiO
Median (IQR) 0.26
0.28 (0.21-0.36)
0.25 (0.21-0.30)
0.57 b
2
(0.21-0.31)
Tidal volume (ml) Inspiratory Expiratory
Median (IQR)
89 (34-157) 91 (34-180)
142 (96-185) 143 (95-198)
36 (20-90) 35 (19-97)
0.002 b 0.001 b
Tube leak (%) (inspiratory-expiratory tidal volume)
Mean ± SD -1.9 ± 8.6 -4.8 ± 10.2 0.3 ± 6.4 0.07 a
Respiratory rate (/min)
Median (IQR) 38 (25-47)
26 (23-35)
43 (38-50) 128 (101-142)
0.02 b 0.27 b
Heart rate (/min)
Median (IQR) 116
113 (99-124)
(103-139)
LOS at moment of measurement (days)
Median (IQR) 2 (1-4)
2 (1-4)
2 (1-4)
0.75 b
Length of measurement (min)
Mean ± SD
30 ± 9.5
34 ± 8
33 ± 11
0.71 a
VCO
2 Deltatrac (ml/min)
Median (IQR) 75.2
108.6 (94-132.9) 111.6 (94.8-135.9)
45.9 (28.3-78.6) 33.4 (12.6-75.3)
<0.001 b
(41.6-116.1)
VCO
2 Servo-I (ml/min)
Median (IQR) 74.0
<0.001 b
(27.5-118.0)
P-values are calculated for differences between the two subgroups using the following tests: a Independent samples t-test, b Mann-Whitney test, c Chi-square test. IQR= Interquartile Range, FiO 2 = Fraction inspired Oxygen, LOS= Length of Stay, SD= Standard Deviation
Correlationbetween Servo-I®-derivedVCO 2 values and IC-derived valueswas excellent [ρ=0.965 (95% CI: 0.935 to 0.981)]. However, despite being a necessary condition for agreement, high correlation does not automatically imply good agreement. Figure 1a shows the Bland-Altman plot for agreement between the two methods for measuring VCO 2 values. Measurements were not comparable with a mean relative difference of 15.6% (p=0.002, one-sample t-test), due to underestimation of VCO 2 values by the Servo-I®. The 95% limits of agreement were wide (-42.4 % and 73.6%). Linear regression of Servo-I®-derived VCO 2 values on IC-derived values showed a regression line with a slope of 1.15 (95% CI: 1.07 to 1.23) and an intercept at -15.66 (95% CI: -23.32 to -8.01). The slope of this regression line was significantly different from 1 (p<0.001), and the intercept was significantly different from 0 (p<0.001), reflecting both proportional and fixed bias.
60